Resources / Regulatory FAQs
Medical Device, IVD and Cosmetics Regulatory FAQs
Clear, practical answers on EU MDR and IVDR, UK MDR, US FDA, Australia TGA, India CDSCO and EU/UK cosmetics regulation, written by RegTrac's regulatory team.
AR, NB, EUDAMED
MHRA and UKRP
510(k), De Novo, PMA
ARTG inclusion
Authorised Agent
EU and UK
About these FAQs
Answers to the regulatory questions manufacturers ask most
This page answers the questions RegTrac’s regulatory team hears most often from medical device, IVD, SaMD and cosmetics manufacturers. Each answer gives the short version first and cites the regulation it comes from, so you can check the source and act quickly.
Topics cover EU Authorised Representatives, Notified Bodies, Class I devices, EUDAMED and technical files under EU MDR and IVDR; UK MDR and the MHRA; US FDA 510(k), De Novo and PMA pathways; ARTG inclusion in Australia; CDSCO registration in India; and EU and UK cosmetics compliance.
Need hands-on help? Explore our MedTech regulatory solutions, Consumer Healthcare services or regulatory training.
Last reviewed: October 2026 by the RegTrac regulatory team. These answers are general guidance, not legal advice. Regulations and timelines change, so always confirm against the current legislation and guidance.
Regulatory dates to know
Key 2025 to 2026 regulatory deadlines
16 June 2025
UK post-market surveillance
Strengthened PMS requirements apply to devices placed on the Great Britain market.
1 October 2025
eSTAR for De Novo
FDA requires De Novo requests to be submitted using the eSTAR template.
2 February 2026
FDA QMSR in force
21 CFR Part 820 now incorporates ISO 13485:2016 by reference.
28 May 2026
EUDAMED mandatory
First four modules mandatory; devices already on the market to be registered by 28 November 2026.
EU MDR Article 11 | IVDR Article 11
EU Authorised Representative FAQs
Non-EU manufacturers must appoint an EU Authorised Representative before placing devices on the EU market. These answers cover who needs one, what the role involves and how liability works under Regulation (EU) 2017/745.
What is an EU Authorised Representative (EC REP)?
An EU Authorised Representative (often called EAR or EC REP) is a natural or legal person established in the EU who holds a written mandate from a non-EU manufacturer to act on its behalf for specified regulatory tasks under EU MDR and IVDR Article 11. The representative verifies key documentation, holds copies for authorities, registers in EUDAMED and is the contact point for Competent Authorities.
Who must appoint an EU Authorised Representative?
Any manufacturer that is not established in an EU Member State and wants to place a medical device or IVD on the EU market. Only one Authorised Representative may be designated per generic device group. Manufacturers established in the EU do not need one.
What are the main responsibilities of an EU Authorised Representative?
Under Article 11(3) the representative must, at a minimum: verify that the EU Declaration of Conformity and technical documentation have been drawn up and that the conformity assessment procedure has been carried out; keep copies of the technical documentation, Declaration of Conformity and certificates available for authorities; comply with registration obligations; provide information and samples to Competent Authorities; forward complaints and incident reports to the manufacturer; cooperate on field safety corrective actions; and terminate the mandate if the manufacturer acts contrary to its obligations.
Is the Authorised Representative legally liable for defective devices?
Yes, in part. Appointing a representative does not remove the manufacturer’s obligations, but under Article 11(5) the Authorised Representative is legally liable for defective devices on the same basis as, and jointly and severally with, the manufacturer where the manufacturer has not complied with its Article 10 obligations. This is why representatives verify documentation carefully before accepting a mandate.
Does an EU Authorised Representative need a PRRC?
Yes. Article 15(6) requires Authorised Representatives to have permanently and continuously at their disposal at least one Person Responsible for Regulatory Compliance (PRRC) with the qualifications required in EU medical device regulatory requirements.
Can one Authorised Representative act for several manufacturers?
Yes. A representative can act for many manufacturers, provided each relationship is covered by its own written mandate that sets out the delegated tasks.
Where does the Authorised Representative appear on the device?
The representative’s name and registered place of business must appear on the label, normally next to the EC REP symbol defined in ISO 15223-1.
Can a manufacturer change its EU Authorised Representative?
Yes. Article 12 requires the change to be defined in an agreement between the manufacturer, the outgoing representative and the incoming representative, covering the end date of the old mandate, the start date of the new one, transfer of documents, labelling transition and the handling of complaints and incidents.
What documentation must the Authorised Representative keep?
A copy of the technical documentation, the EU Declaration of Conformity and any relevant certificates (including amendments and supplements), available to Competent Authorities for the retention period set out in Article 10(8): at least 10 years after the last device covered by the Declaration of Conformity is placed on the market, or 15 years for implantable devices.
EU MDR Chapter IV | UK MDR 2002
Notified Bodies and Approved Bodies FAQs
Most devices above standard Class I need an independent conformity assessment. These answers explain how Notified Bodies (EU) and UK Approved Bodies work and how to choose one.
What is a Notified Body?
A conformity assessment body designated by an EU Member State authority, and notified to the European Commission, to assess medical devices under EU MDR 2017/745 or IVDs under EU IVDR 2017/746. Its designation sets out the device types and procedures it can assess.
What does a Notified Body do during conformity assessment?
Depending on device class and the conformity assessment route, it audits the manufacturer’s quality management system, reviews technical documentation (including clinical or performance evidence), may test or sample devices, issues certificates, and carries out surveillance and unannounced audits.
What is the difference between a Notified Body and a UK Approved Body?
Notified Bodies are designated under EU legislation and listed in the European Commission’s NANDO database. UK Approved Bodies are designated by the MHRA under the UK Medical Devices Regulations 2002 (as amended) and issue UKCA certificates for the Great Britain market.
Is a Notified Body the same as a Competent Authority?
No. A Competent Authority is the government regulator in each country (for example the HPRA in Ireland or BfArM in Germany; the MHRA in the UK). It designates and monitors conformity assessment bodies, handles vigilance and enforces the law. A Notified Body is an independent organisation that performs conformity assessment.
Can any Notified Body assess any device?
No. Each Notified Body has a designated scope covering specific device codes, technologies and risk classes. Check the NANDO listing to confirm the scope covers your device before you apply, and allow for capacity and lead times.
What is the NANDO database?
NANDO (New Approach Notified and Designated Organisations) is the European Commission database that lists Notified Bodies designated under EU harmonisation legislation, including the MDR and IVDR, together with their scope.
Which devices need Notified Body involvement under EU MDR?
Class IIa, IIb and III devices, and Class I devices that are sterile (Is), have a measuring function (Im) or are reusable surgical instruments (Ir). Standard Class I devices are self-declared by the manufacturer.
EU MDR Annex VIII | Article 52(7)
EU Class I Medical Devices FAQs
Class I is the lowest-risk class under EU MDR, but manufacturers still carry full documentation, clinical evaluation and post-market obligations.
What are Class I medical devices under EU MDR?
Low-risk devices classified under the rules in EU MDR Annex VIII, typically non-invasive or with limited body contact. Examples include manual wheelchairs, non-sterile dressings, tongue depressors and hospital beds.
What are the Class I subclasses (Is, Im and Ir)?
Class Is covers devices placed on the market in sterile condition, Class Im covers devices with a measuring function and Class Ir covers reusable surgical instruments. For these, a Notified Body is involved only for the aspects relating to sterility, metrology or reprocessing.
Can Class I manufacturers self-certify?
Yes, for standard Class I devices. The manufacturer completes the technical documentation, draws up the EU Declaration of Conformity and affixes the CE mark without a Notified Body. Is, Im and Ir devices need a Notified Body certificate for the relevant aspects.
What is the conformity assessment process for a Class I device?
Confirm classification under Annex VIII; establish a proportionate QMS (Article 10(9)); prepare technical documentation under Annexes II and III; apply risk management to ISO 14971; complete a clinical evaluation; set up post-market surveillance and vigilance; involve a Notified Body if Is, Im or Ir; sign the EU Declaration of Conformity; affix the CE mark; and register the device in EUDAMED.
What must appear on a Class I device label?
Requirements are in EU MDR Annex I, Section 23. They include the device name or trade name, manufacturer name and address, Authorised Representative and importer details where applicable, UDI carrier, lot or serial number, sterile status where relevant, storage conditions, warnings and the CE mark (with the Notified Body number for Is, Im and Ir).
Does a Class I device need a clinical investigation?
Not usually. A clinical evaluation under Article 61 is always required, but for low-risk devices it is commonly based on existing clinical data and literature. A clinical investigation is needed only if existing data cannot demonstrate conformity with the relevant safety and performance requirements.
What are the ongoing post-market obligations for Class I manufacturers?
Keep the technical documentation up to date, run a post-market surveillance system and prepare a PMS report (Article 85) updated when necessary, report serious incidents and field safety corrective actions through vigilance (Articles 87 to 89), and maintain EUDAMED registrations.
EU MDR Articles 33 and 34 | Regulation (EU) 2024/1860
EUDAMED FAQs
Use of the first four EUDAMED modules became mandatory on 28 May 2026, with a transition period for devices already on the market.
What is EUDAMED?
The European Database on Medical Devices, set up under EU MDR and IVDR to collect and share information on economic operators, devices (UDI), certificates, clinical investigations, vigilance and market surveillance. Access it on the European Commission EUDAMED site.
When did EUDAMED become mandatory?
Following the gradual roll-out introduced by Regulation (EU) 2024/1860, use of the first four modules (actor registration, UDI/device registration, Notified Bodies and certificates, and market surveillance) became mandatory from 28 May 2026. The remaining modules become mandatory once the Commission confirms their functionality.
Is there a transition period for devices already on the market?
Yes. Devices placed on the market before 28 May 2026 that remain on the market generally have until 28 November 2026 to be registered. Check the current Commission guidance for your device type, as different rules apply to legacy devices and certificates.
Who must register in EUDAMED?
Manufacturers, Authorised Representatives, importers and system or procedure pack producers register as actors and receive a Single Registration Number (SRN). Manufacturers then register their devices with UDI data.
What are the six EUDAMED modules?
(1) Actor registration, (2) UDI and device registration, (3) Notified Bodies and certificates, (4) clinical investigations and performance studies, (5) vigilance and post-market surveillance, and (6) market surveillance.
How does UDI relate to EUDAMED?
The Unique Device Identifier (UDI) identifies a device model (UDI-DI) and production unit (UDI-PI). The Basic UDI-DI and UDI-DI data are submitted to EUDAMED, which links the device to its manufacturer, certificates and vigilance data.
What happens if an operator does not register in EUDAMED?
Where registration is mandatory, the device cannot lawfully be placed on the market, and Competent Authorities can take enforcement action.
How often should EUDAMED records be updated?
Whenever relevant information changes, such as company details, device data or certificate status. Actor data must be confirmed periodically, and changes should be made within the timelines set in the MDR.
What is the EUDAMED playground?
A test environment where economic operators can practise registrations and data submissions without affecting the live database. It is useful for training teams and testing bulk upload (XML/M2M) set-ups.
EU MDR Annexes II and III
Medical Device Technical File (MDTF) FAQs
Technical documentation is the evidence behind every CE or UKCA mark. These answers cover content, review, updates and retention.
Related: Class I technical file service | Medical writing | Risk management
What is a medical device technical file?
The technical documentation a manufacturer prepares to demonstrate that a device meets the applicable regulatory requirements. Under EU MDR it is defined in Annex II (technical documentation) and Annex III (post-market surveillance documentation). It is required for every device class in the EU and UK.
What must an EU MDR technical file contain?
Device description and specification (including variants and accessories); information supplied by the manufacturer (labels and IFU); design and manufacturing information; General Safety and Performance Requirements (GSPR) checklist; benefit-risk analysis and risk management; product verification and validation (including clinical evaluation); and the post-market surveillance plan and reports.
Who prepares and who reviews the technical file?
The manufacturer prepares it. For devices that need third-party assessment, the Notified Body (EU) or UK Approved Body (GB) reviews it as part of conformity assessment. Standard Class I files are not reviewed before market entry but must be available to authorities on request.
Do the Authorised Representative or UK Responsible Person need access to the technical file?
Yes. Both must verify that the technical documentation exists and keep a copy, or have access to it, so they can provide it to the Competent Authority or MHRA on request. See our guide on UK Responsible Person responsibilities.
When does the technical file need to be updated?
Whenever there is a change that could affect safety, performance or conformity, such as design, materials, suppliers, intended purpose or labelling changes, and as post-market surveillance, clinical and risk management data are updated. Significant changes to certified devices must be notified to the Notified Body before implementation.
How long must the technical file be retained?
Under EU MDR Article 10(8), for at least 10 years after the last device covered by the EU Declaration of Conformity has been placed on the market, or at least 15 years for implantable devices.
Where can I find official guidance on technical documentation?
Start with EU MDR Annexes II and III and the relevant MDCG guidance documents for the EU, and the UK Medical Devices Regulations 2002 (as amended) with MHRA guidance for Great Britain.
UK MDR 2002 | MHRA
UK Medical Device Regulation (MHRA) FAQs
Great Britain regulates devices under the UK Medical Devices Regulations 2002 (as amended), overseen by the MHRA. Northern Ireland follows EU MDR and IVDR.
Who regulates medical devices in the UK?
The Medicines and Healthcare products Regulatory Agency (MHRA). Devices placed on the Great Britain market must be registered with the MHRA before they are placed on the market.
Do non-UK manufacturers need a UK Responsible Person?
Yes. A manufacturer based outside the UK must appoint a UK Responsible Person (UKRP) established in the UK to register devices with the MHRA and act on its behalf for specified tasks. The UKRP’s details must appear on the label or packaging where required.
Are CE-marked devices still accepted in Great Britain?
Yes, for now. Current legislation allows CE-marked devices to be placed on the GB market until 30 June 2028 (MDD and AIMDD compliant devices) or 30 June 2030 (EU MDR and IVDR compliant devices), subject to conditions. In 2026 the MHRA consulted on indefinite recognition of CE marking, so check the latest MHRA position before planning your UKCA route.
What is the difference between UKCA and CE marking?
The UKCA mark shows conformity with UK MDR 2002 and is assessed by a UK Approved Body where third-party assessment is needed. The CE mark shows conformity with EU legislation and is assessed by an EU Notified Body. Northern Ireland requires CE marking (with UKNI marking where a UK Notified Body is used).
What are the UK post-market surveillance requirements?
Strengthened post-market surveillance requirements for Great Britain apply from 16 June 2025. Manufacturers need a PMS system and plan, periodic reporting proportionate to device class, and serious incident reporting to the MHRA within defined timelines.
21 CFR Parts 807, 812, 814 | FD&C Act
US FDA Device Submissions FAQs
The FDA regulates devices through risk-based classification and several premarket pathways. These answers summarise who regulates, how devices are classified and which submission applies.
Who regulates medical devices in the USA?
The US Food and Drug Administration (FDA), mainly through the Center for Devices and Radiological Health (CDRH). The Center for Biologics Evaluation and Research (CBER) regulates certain devices, such as those used in blood collection and processing and some IVDs for blood screening.
How are medical devices classified in the USA?
Class I (low risk, general controls; most are 510(k)-exempt), Class II (moderate risk, general and special controls; most need a 510(k)) and Class III (high risk; usually need Premarket Approval). Classification is set by the device’s product code and classification regulation in 21 CFR Parts 862 to 892.
What are the main FDA premarket pathways?
Premarket Notification 510(k), De Novo classification, Premarket Approval (PMA) and Humanitarian Device Exemption (HDE). An Investigational Device Exemption (IDE) allows clinical studies before marketing authorisation.
What is an Investigational Device Exemption (IDE)?
An IDE under 21 CFR Part 812 allows an investigational device to be used in a clinical study to collect safety and effectiveness data. Significant-risk studies need FDA IDE approval and IRB approval; non-significant-risk studies can proceed with IRB approval under abbreviated requirements.
What is Premarket Approval (PMA)?
The FDA’s most stringent review, used for most Class III devices (21 CFR Part 814). The manufacturer must provide valid scientific evidence, usually including clinical data, giving reasonable assurance of safety and effectiveness.
What is a Humanitarian Device Exemption (HDE)?
A pathway for Humanitarian Use Devices intended for diseases or conditions affecting not more than 8,000 people in the US per year. It requires evidence of safety and probable benefit rather than full effectiveness data.
What must appear on a US device label?
Labelling requirements are in 21 CFR Part 801 and include the device name, manufacturer, packer or distributor name and place of business, adequate directions for use (or prescription device statements), warnings, and the UDI unless exempt (21 CFR Part 801 Subpart B).
Do foreign manufacturers need a US Agent?
Yes. Foreign establishments that register with the FDA must designate a US Agent under 21 CFR 807.40. The agent is the FDA’s point of contact for the establishment.
21 CFR 807 Subpart E | eSTAR
FDA 510(k), Q-Sub and De Novo FAQs
The 510(k) is the most common US route for Class II devices. Since October 2023, 510(k) submissions must use the FDA eSTAR template, and De Novo requests followed from 1 October 2025.
What is a 510(k) Premarket Notification?
A submission showing that a device is substantially equivalent to a legally marketed predicate device in intended use and technological characteristics, or that differences do not raise new questions of safety and effectiveness. Most Class II and some Class I devices need one unless exempt.
What are the three types of 510(k)?
Traditional 510(k) for most new devices; Special 510(k) for certain changes to the manufacturer’s own cleared device where design control activities can be summarised; and Abbreviated 510(k), which relies on FDA guidance, special controls or recognised consensus standards.
When is a 510(k) required?
When introducing a device to the US market for the first time, when changing a marketed device in a way that could significantly affect safety or effectiveness, or when marketing it for a new or different intended use.
When is a 510(k) not required?
For 510(k)-exempt devices; pre-amendment devices legally marketed before 28 May 1976 and not significantly changed; components sold to other manufacturers for further processing; and distributors relabelling or repackaging another firm’s cleared device without changes (subject to conditions).
How are 510(k) and De Novo submissions filed?
Electronically using the eSTAR interactive template through the CDRH Portal. eSTAR has been mandatory for 510(k)s since 1 October 2023 and for De Novo requests since 1 October 2025.
What is a Pre-Submission (Q-Sub) and when should I use one?
A Pre-Submission is a request for written FDA feedback, with an optional meeting, before a formal submission. It is most valuable for novel devices, unclear predicates, De Novo requests, and testing or clinical study protocols.
What is the De Novo classification process?
A risk-based pathway for novel low-to-moderate risk devices with no legally marketed predicate. A manufacturer can request De Novo classification directly or after a Not Substantially Equivalent (NSE) 510(k) decision.
What happens after FDA grants a De Novo request?
The device is classified into Class I or II, a new classification regulation and product code are created, and the device can then serve as a predicate for future 510(k) submissions.
Therapeutic Goods Act 1989 | MD Regulations 2002
Australia (TGA) FAQs
Medical devices must be included in the Australian Register of Therapeutic Goods (ARTG) before supply in Australia. See the TGA overview of device regulation.
Related: Global registrations | All RegTrac solutions
Who regulates medical devices in Australia?
The Therapeutic Goods Administration (TGA), part of the Australian Government Department of Health, Disability and Ageing. Devices must be included in the ARTG before they can be supplied in Australia.
How does the TGA define a medical device?
Broadly, a product intended for diagnosis, prevention, monitoring, treatment or alleviation of disease or injury, or for investigation or modification of anatomy or a physiological process, that does not achieve its principal intended action by pharmacological, immunological or metabolic means (Section 41BD of the Therapeutic Goods Act 1989).
How are devices classified in Australia?
Medical devices are classified as Class I, Class I (sterile), Class I (measuring), Class IIa, Class IIb, Class III or Active Implantable Medical Devices (AIMD). IVDs are classified as Class 1 to 4. Classification depends on intended purpose and risk.
How do overseas manufacturers get ARTG inclusion?
Appoint an Australian sponsor, who submits the application. The sponsor obtains manufacturer evidence (TGA conformity assessment or acceptable overseas evidence such as EU Notified Body certificates), prepares the Declaration of Conformity, submits the application in the TGA Business Services (TBS) portal with the GMDN term, responds to any application audit and receives the ARTG entry.
What documents are typically needed for ARTG inclusion?
Manufacturer evidence (such as a CE certificate or TGA conformity assessment certificate), the Australian Declaration of Conformity, technical documentation available for audit, labelling and IFU, and the sponsor agreement with the manufacturer.
What must appear on a device label in Australia?
Information required by the Essential Principles (Schedule 1, Essential Principle 13) includes the manufacturer’s name and address, the Australian sponsor’s name and address, device name and intended purpose, batch or serial number, expiry date where relevant, sterility information, warnings and instructions for use.
What is a STED and what does it contain?
The Summary Technical Documentation (STED) format, based on GHTF/IMDRF guidance, summarises device description, Essential Principles checklist, risk management summary, design and manufacturing information, clinical evidence, verification and validation, and labelling. The TGA may request it during an application audit.
Medical Devices Rules 2017
India (CDSCO) FAQs
Foreign manufacturers supply India through an Indian Authorised Agent holding a CDSCO import licence. Read our guide on how to register medical devices in India.
Who regulates medical devices in India?
The Central Drugs Standard Control Organisation (CDSCO), headed by the Drugs Controller General of India (DCGI), under the Ministry of Health and Family Welfare. Devices are regulated under the Medical Devices Rules 2017.
How are devices classified in India?
Class A (low risk), Class B (low-moderate risk), Class C (moderate-high risk) and Class D (high risk), using the classification rules in the First Schedule of the Medical Devices Rules 2017.
Can a foreign manufacturer sell devices in India?
Yes, through an Indian Authorised Agent who applies for the import licence on the manufacturer’s behalf. The manufacturer cannot hold the import licence directly without a local entity.
What is an India Authorised Agent (IAA)?
A person or company established in India and authorised by a foreign manufacturer, through a Power of Attorney, to handle CDSCO applications, regulatory correspondence and post-market obligations on its behalf.
Who can act as an India Authorised Agent?
An entity in India that holds a valid wholesale (sale or distribution) licence or a manufacturing licence, and that can meet the regulatory and post-market responsibilities set out in the Medical Devices Rules 2017.
What are the responsibilities of an India Authorised Agent?
Submitting import licence applications to CDSCO; responding to queries; keeping technical documentation available; handling complaints, adverse event reporting and field safety corrective actions; and acting as the local contact for CDSCO.
Does a manufacturer based in India need an Authorised Agent?
No. Indian manufacturers apply for a manufacturing licence themselves.
Can a manufacturer change its India Authorised Agent?
Yes, but the import licence is held by the agent, so a change usually requires a new or transferred licence. Plan the transition so that supply and post-market obligations are not interrupted.
Regulation (EC) 1223/2009 | UK SI 2019/696
EU and UK Cosmetics FAQs
Cosmetics in the EU and Great Britain need a Responsible Person, a Product Information File and a pre-market notification. Since Brexit the two markets are notified separately.
Related: Consumer Healthcare services | US MoCRA compliance
What is a cosmetic product under EU and UK law?
Any substance or mixture intended to be placed in contact with the external parts of the human body (skin, hair, nails, lips, external genital organs) or with the teeth and oral mucous membranes, with a view exclusively or mainly to cleaning, perfuming, changing appearance, protecting, keeping in good condition or correcting body odours (Article 2 of Regulation (EC) 1223/2009).
Who is the Responsible Person for cosmetics?
The legal or natural person who ensures each product complies with the regulation before it is placed on the market. For the EU the Responsible Person must be established in the EU; for Great Britain, in the UK. Products sold in both markets need a Responsible Person for each.
What are the Responsible Person's key duties?
Holding the Product Information File (PIF); ensuring a Cosmetic Product Safety Report (CPSR) is completed; checking labelling and claims; notifying products before launch; monitoring and reporting serious undesirable effects; and acting as the contact point for authorities.
What is the Product Information File (PIF)?
A file kept for 10 years after the last batch is placed on the market. It contains the product description, the Cosmetic Product Safety Report, manufacturing method and GMP statement, proof of claimed effects and data on any animal testing.
What is the Cosmetic Product Safety Report (CPSR)?
The safety assessment required by Annex I of Regulation 1223/2009. Part A sets out the safety information (formula, specifications, exposure, toxicology, undesirable effects) and Part B contains the safety assessor’s conclusion, labelled warnings and reasoning. Part B must be signed by a qualified safety assessor.
How are cosmetics notified in the EU and UK?
In the EU through the Cosmetic Products Notification Portal (CPNP). In Great Britain through the Submit Cosmetic Product Notifications (SCPN) service run by the Office for Product Safety and Standards (OPSS). Notification is required before the product is placed on the market.
Is animal testing allowed for cosmetics?
No. The EU and UK prohibit animal testing of finished cosmetic products and ingredients for cosmetic purposes, and the EU also bans marketing cosmetics tested on animals to meet the regulation’s requirements.
What GMP standard applies to cosmetics?
ISO 22716 (Cosmetics, Good Manufacturing Practices) is the harmonised standard used to demonstrate GMP compliance in the EU and recognised in the UK.
What must appear on a cosmetic label?
Under Article 19: Responsible Person name and address, country of origin for imported products, nominal content, date of minimum durability or period after opening, precautions, batch number, function (if not clear from presentation) and the ingredient list using INCI names.
How are cosmetic claims regulated?
Claims must meet the common criteria in Regulation (EU) 655/2013: legal compliance, truthfulness, evidential support, honesty, fairness and informed decision-making. Medicinal or therapeutic claims can move a product out of cosmetics law into medicines or medical device regulation.